APPLICATION:
Natural Gas Wellhead Trace Oxygen Analysis and Shut-In |
|
DESCRIPTION:
Local gathering company requires natural gas received from producer
to have less than 10 ppm (parts per million) oxygen. Higher levels are
indicative of a leak, which are much more prevalent in this low-pressure
field. Oxygen, combined with the moisture inherent in the gas stream,
creates a dangerous corrosion hazard in the gatherers high-pressure
system.
SOLUTION: Winn-Marion Barber, LLP provides turnkey
supply and installation of this AMI analyzer and Jamesbury “slam-valve”
package. When the oxygen level reaches a point of 10 ppm or higher,
a pulse is sent to the valve to slam it shut. This gives the producer
the opportunity to inspect his well and compression to find the leak.
Otherwise the high oxygen reading would be seen by the gathering company
and the producer’s well would be shut-in by the gatherer. When
this happens, the downtime before the well can be brought on-line may
be days, resulting in great loss of production and income.
|